1586A
Users Manual
7-16
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 704
Error reading
calibration folder
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 705
Error reading
calibration history
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 707
Error reading
calibration constants
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 708
Error writing
calibration constants
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 709
Error restoring LAN
settings
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 710
Error writing LAN
settings
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 711
Error setting MAC
address
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 712
Error storing MAC
address
Cause:
Internal flash is damaged.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 800
Serial framing error
Cause:
RS232 framing error
Solution:
Make sure that the PC and Instrument settings are
the same.
Error 821
LAN port out of range
Cause:
The LAN port < 1024 or > 65535.
Solution:
Check the port address and try again.
Error 822
Error opening LAN
port
Cause:
Error while opening LAN port.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 824
LAN port receive
error
Cause:
Error while reading LAN port.
Solution:
Check the LAN connection.
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...
Страница 114: ...1586A Users Manual 6 2...
Страница 118: ...1586A Users Manual 6 6...
Страница 119: ...7 1 Chapter 7 Error Messages and Troubleshooting Title Page Introduction 7 3 Error Messages 7 3 Troubleshooting 7 20...
Страница 120: ...1586A Users Manual 7 2...
Страница 140: ...1586A Users Manual 7 22 Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it...