Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 502
Temperature unit
locked
Cause:
UNIT:TEMPerature command is received while the
temperature unit is locked to °C.
Solution:
Contact Fluke. See Chapter 1.
Error 527
Operations not
allowed while busy
Cause:
Operation cannot be performed while system is busy.
Solution:
Use *RST command to reset the Instrument to the
factory default state.
Error 528
Channel in use for
3-wire or 4-wire
Cause:
The channel is unavailable or reserved for 3-wire or
4-wire use.
Solution:
Reconfigure the channel that reserved this sense
channel for use. See Chapter 3.
Error 530
Configuration file
corrupt
Cause:
The test setup is found corrupted while loading.
Solution:
Cycle the power and try to reload the setup file. If
the error occurs again, load a different file to see if the the
single file is corrupted or if the Instrument is malfunctioning. If
the error happens with a different file, contact Fluke (see
Chapter 1).
Error 580
(STA) begin not
called
Cause:
Invalid accessing to state.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 590
Cannot get DHCP IP
address
Cause:
Get an IP address from DHCP server failed.
Solution:
Make sure a DHCP server is running in your
network.
Error 600
(DB) begin not called
Cause:
Invalid accessing to database.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 601
(DB) Error initializing
data buffer
Cause:
Database is not initialized.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 602
(DB) Data buffer
overflow
Cause:
Database buffer overflow.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 603
(DB) Data not
available
Cause:
Database empty.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...
Страница 114: ...1586A Users Manual 6 2...
Страница 118: ...1586A Users Manual 6 6...
Страница 119: ...7 1 Chapter 7 Error Messages and Troubleshooting Title Page Introduction 7 3 Error Messages 7 3 Troubleshooting 7 20...
Страница 120: ...1586A Users Manual 7 2...
Страница 140: ...1586A Users Manual 7 22 Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it...