Error Messages and Troubleshooting
Error Messages
7
7-7
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 207
Failed to reset relay
on slot 1
Cause:
The relay does not reset when the channel is
switched.
Solution:
Cycle the power and try again. If the error occurs
again, the relay card needs serviced, contact Fluke (see
Chapter 1).
Error 208
Failed to reset relay
on slot 2
Error 209
Failed to reset relay
on slot 3
Error 210
Error reading cold
junction temperature
(Slot 1)
Cause:
The CJC temperature cannot be read from external
module of slot 1.
Solution:
Cycle the power. If the error occurs again, the relay
card needs serviced, contact Fluke (see Chapter 1).
Error 211
Error reading cold
junction temperature
(Slot 2)
Cause:
The CJC temperature cannot be read from external
module of slot 2.
Solution:
Cycle the power. If the error occurs again, the relay
card needs serviced, contact Fluke (see Chapter 1).
Error 260
Error copying
configure to ChXXX
Cause:
The copy function failed.
Solution:
Check the channel list and make sure all selected
channel are set up correctly.
Error 300
Invalid procedure
Cause:
Wrong calibration procedure is specified in
CALibrate:STARt command or MAIN in CALibrate:STARt
command could be misspelled.
Solution:
Check the parameter value and try again.
Error 301
Invalid procedure
step
Cause:
Wrong or misspelled calibration step is specified in
CALibrate:STARt command.
Solution:
Check the parameter value and try again.
Error 302
Step change not
allowed while busy
Cause:
Invalid CALibrate:STARt was received while a
calibrate operation is in progress.
Solution:
Send CALibrate:ABORt to terminate the active
calibration procedure.
Error 303
Invalid starting point
Cause:
The step specified in CALibrate:STARt command
should be the first point of a function or a range, it is invalid to
start a calibration procedure from an intermediate step.
Solution:
Check the start point and try again.
Error 304
Invalid reference unit
Cause:
CALibrate:NEXT command was received with an
unexpected unit.
Solution:
Check the parameter value and try again.
Error 305
Entered value out of
range
Cause:
CALibrate:NEXT command was received, but the
reference value is out of range.
Solution:
Check the parameter value and try again.
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...
Страница 114: ...1586A Users Manual 6 2...
Страница 118: ...1586A Users Manual 6 6...
Страница 119: ...7 1 Chapter 7 Error Messages and Troubleshooting Title Page Introduction 7 3 Error Messages 7 3 Troubleshooting 7 20...
Страница 120: ...1586A Users Manual 7 2...
Страница 140: ...1586A Users Manual 7 22 Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it...