1586A
Users Manual
7-18
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error 910
File locations full
Cause:
The number of setup file reaches the limit (99).
Solution:
Remove some setup files.
Error 1302
Bad parameter count
Cause:
The number of parameter does not match the
command.
Solution:
Check the parameter and try again.
Error 1305
Illegal parameter unit
Cause:
Illegal parameter unit was specified in the command.
The valid parameter unit is UV, MV, V, KV, UA, MA, A, OHM,
KOHM, MOHM, GOHM, TOHM, HZ, KHZ, MHZ, GHZ, CEL,
FAR.
Solution:
Check the parameter and try again.
Error 1313
Operation only
allowed for service
Cause:
The service only command was received while
service mode is not activated.
Solution:
Enter service mode and try the operation again.
Error 1315
Invalid device trigger
Cause:
An invalid bus trigger *TRG was received.
Solution:
Input Ctrl+T when input buffer is not empty.
Error 1317
Communication input
buffer overrun
Cause:
Input buffer overrun for communication.
Solution:
Cycle the power. If the error occurs again, contact
Fluke (see Chapter 1).
Error 1320
Invalid binary data
Cause:
An invalid binary data parameter was found in the
command.
Solution:
Check the parameter and try again.
Error 1323
Invalid decimal data
Cause:
An invalid decimal data parameter was found in the
command.
Solution:
Check the parameter and try again.
Error 1326
Invalid hexadecimal
data
Cause:
An invalid hexadecimal data parameter was found in
the command.
Solution:
Check the parameter and try again.
Error 1328
Invalid octal data
Cause:
An invalid octal data parameter was found in the
command.
Solution:
Check the parameter and try again.
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...
Страница 114: ...1586A Users Manual 6 2...
Страница 118: ...1586A Users Manual 6 6...
Страница 119: ...7 1 Chapter 7 Error Messages and Troubleshooting Title Page Introduction 7 3 Error Messages 7 3 Troubleshooting 7 20...
Страница 120: ...1586A Users Manual 7 2...
Страница 140: ...1586A Users Manual 7 22 Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it...