1586A
Users Manual
1-4
•
Data Storage -
Store up to 20 MB of data and channel setup files directly to the
internal non-volatile memory or an external USB drive. Data can also be transferred
to a PC with a USB drive or LAN TCP/IP interface at the rear of the unit with SCPI
interface command sets.
•
Alarms –
Each channel can be assigned two independent alarms to indicate when
either an upper (HI) or lower (LO) range has been exceeded. The alarms can be
configured to output a digital signal from the rear-panel alarm output for control of
external devices.
•
Digital I/O (DIO) –
The Product is equipped with a digital 8-bit transistor-transistor
logic (TTL) port that can sense and output. When the DIO channel is included in the
scan list, the value of this port is recorded in the scan data record on each scan with a
value range of 0 to 255 based on the port state when it is read.
•
Totalizer –
The Product is equipped with a unidirectional, resettable totalizer with an
input count capability of 1,048,575 (20 bits). Counter increment is accomplished
through a change to a digital signal or contact closure to the totalizer input terminals
on the rear of the Instrument.
•
Remote Operation –
Remotely operate the Product with remote SCPI commands
from application software over a rear-panel USB or LAN TCP/IP connection.
•
Automated Sensor Test
–
Automatically sets the temperature of a Fluke
Calibration dry-well calibrator or temperature bath connected to the rear-panel
RS-232 port. Monitors temperature stability, and measures the readings of
temperature sensors and a reference thermometer when the temperature is stable.
•
DAQ-STAQ Accessory
–
Multiplexer module provides quick and convenient
connection of temperature sensors with Fluke's patented spring-loaded DWF
terminals.
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...
Страница 114: ...1586A Users Manual 6 2...
Страница 118: ...1586A Users Manual 6 6...
Страница 119: ...7 1 Chapter 7 Error Messages and Troubleshooting Title Page Introduction 7 3 Error Messages 7 3 Troubleshooting 7 20...
Страница 120: ...1586A Users Manual 7 2...
Страница 140: ...1586A Users Manual 7 22 Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it...