1586A
Users Manual
7-4
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error -161
Invalid block data
Cause:
A block data element was expected, but was invalid.
For example, an END message was received before the
length was satisfied.
Solution:
Check the command and try again.
Error -151
Invalid string data
Cause:
An invalid character string was received.
Solution:
Check to see if the character string in is enclosed
in quotation marks and verify that the string contains valid
ASCII characters.
Error -144
Character data too
long
Cause:
The character data element contains more than
250 characters.
Solution:
Check the parameter value and try again.
Error -123
Exponent too large
Cause:
A numeric parameter was found whose exponent was
too large for this command.
Solution:
Check the parameter value and try again.
Error -113
Undefined header
Cause:
A command was received that is not valid for this
instrument.
Solution:
The command entered could be misspelled or not
a valid command. If the shortened form of a command is
used, remember that it can contain up to four letters. Also
check for duplicate colons.
Error -109
Missing parameter
Cause:
Fewer parameters were received than expected for
this command.
Solution:
One or more parameters that are required for this
command have been omitted. Check the parameter value
and try again.
Error -108
Parameter not
allowed
Cause:
More parameters were received than expected for
this command.
Solution:
An extra parameter could have been added to a
command that does not require a parameter. Check the
parameter value and try again.
Error -104
Data type error
Cause:
Invalid data type was found in the command string.
Solution:
Check the parameter value and try again.
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...
Страница 114: ...1586A Users Manual 6 2...
Страница 118: ...1586A Users Manual 6 6...
Страница 119: ...7 1 Chapter 7 Error Messages and Troubleshooting Title Page Introduction 7 3 Error Messages 7 3 Troubleshooting 7 20...
Страница 120: ...1586A Users Manual 7 2...
Страница 140: ...1586A Users Manual 7 22 Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it...