Error Messages and Troubleshooting
Error Messages
7
7-5
Table 7-1. Error Messages (cont.)
Error Code
Error Message
Causes and Solutions
Error -102
Syntax error
Cause:
Invalid syntax was found in the command string.
Solution:
Check the parameter value and try again.
Error 0
No error
Cause:
Error queue is empty. No action necessary.
Error 100
Acknowledgement
queue full
Cause:
GX acknowledgement queue overflow.
Note
GX is the abbreviation for Guard Crossing - which is the
communication link between the inguard and outguard.
Solution:
Instrument will restart the GX.
Error 101
Inguard not
responding (recv)
Cause:
Inguard not responding (receive).
Solution:
Instrument will restart the GX.
Error 102
Lost sync with
inguard
Cause:
Lost sync with inguard.
Solution:
Instrument will restart the GX.
Error 103
Inguard not
responding (send)
Cause:
Inguard not responding (send).
Solution:
Instrument will restart the GX.
Error 104
Oversize packet
received
Cause:
Oversize packet received.
Solution:
The packet will be ignored automatically.
Error 105
GX CRC error
Cause:
CRC error detected.
Solution:
No user action necessary. The packet will be
ignored automatically.
Error 106
Wrong ACK number
Cause:
An ACK packet was received with unexpected ACK
number.
Solution:
No user action necessary. The Instrument will
restart the GX.
Error 107
Info packet received;
link not active
Cause:
An INFO packet was received, but the
communication has not been setup.
Solution:
No user action necessary. The packet will be
ignored automatically.
Error 108
Unknown control byte
Cause:
A packet was received with unknown control byte.
Solution:
No user action necessary. Instrument will restart
the GX.
Содержание 1586A
Страница 6: ...1586A Users Manual iv Error Messages 7 3 Troubleshooting 7 20...
Страница 8: ...1586A Users Manual vi...
Страница 10: ...1586A Users Manual viii...
Страница 12: ...1586A Users Manual 1 2 General 1 22 1586 2586 High Capacity Input Module Specifications 1 22 General 1 22...
Страница 34: ...1586A Users Manual 2 2...
Страница 42: ...1586A Users Manual 2 10 2 3 4 1 hcn018 eps Figure 2 6 Relay Card Installation...
Страница 44: ...1586A Users Manual 2 12 hcn032 eps Figure 2 7 DAQ STAQ Multiplexer Installation...
Страница 46: ...1586A Users Manual 2 14...
Страница 48: ...1586A Users Manual 3 2...
Страница 82: ...1586A Users Manual 3 36...
Страница 84: ...1586A Users Manual 4 2...
Страница 108: ...1586A Users Manual 5 2...
Страница 114: ...1586A Users Manual 6 2...
Страница 118: ...1586A Users Manual 6 6...
Страница 119: ...7 1 Chapter 7 Error Messages and Troubleshooting Title Page Introduction 7 3 Error Messages 7 3 Troubleshooting 7 20...
Страница 120: ...1586A Users Manual 7 2...
Страница 140: ...1586A Users Manual 7 22 Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it...