Conversions for each of the thermocouple types accept optional calibration
data. This can be used to improve the measurement accuracy. CAL PTS indi-
cates the number of calibration points used, up to three. T
n
is the temperature
of the point. ADJ
n
is the temperature deviation from the reference function at
the point. The temperature measurement will be adjusted by this amount when
measuring at this temperature. A polynomial interpolation function is used for
measurements between calibration points. If calibration data is not available or
is not to be used, the ADJ
n
parameters should all be set to 0.0. This will cause
the temperature to be calculated according to the standard reference functions.
Some thermocouple conversions (types R, S, and gold-platinum) also accept
polynomial calibration coefficients. This produces a temperature-dependent ad-
justment to the voltage according to the polynomial function:
Δ
Δ
Δ
E t
c t
c t
s
s
s
( [
])[
]
°
=
+
C
V
μ
1
2
2
The probe parameters DC1 and DC2 are used to set the values of
Δ
c
1
and
Δ
c
2
respectively. If no calibration coefficients are available or if only the standard
reference function is to be used, DC1 and DC2 should be set to 0.0. If these co-
efficients are used, the ADJn parameters described above must be set to 0.
5.2.1.11
Thermocouple Table Conversion
EDIT TCS 03
PROBE SER#: 1
CONVERSION: TC-TABLE
CJC: INTERNAL
CJC TEMP: 0.0
TABLE PTS: 10
T1(C): -200.0
$
The thermocouple table conversion allows calculation of temperature by inter-
polating from a table. You can enter the temperatures (in °C) and voltages (in
V) from a reference or calibration table for one to ten points. The number of
points is specified by TABLE PTS. To calculate temperature, a polynomial in-
terpolation is done using up to four of the points in the table closest to the given
temperature or voltage. The zero point (0 V at 0°C) is automatically included in
the table whether it is explicitly specified or not.
1560 Thermometer Readout
User’s Guide
42
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...