1560 Thermometer Readout
User’s Guide
164
GPIB (IEEE-488) interface
Capability
IEEE-488.2, 1992
SH1, complete source handshake capability
AH1, complete acceptor handshake capability
T6, basic talker, serial poll, no talk-only capability, unaddress if MLA
TE0, no extended-talk capability
L4, basic listener, no listen-only capability, unaddress if MTA
LE0, no extended-listen capability
SR1, complete serial poll capability
RL0, no remote-local capability
PP0, no parallel poll capability
DC1, complete device clear capability
DT0, no device trigger capability
C0, no control capability
E2, three-state drivers
Parallel (Centronics) printer interface
Connector
25-pin subminiature D
Serial (RS-232) interface
Connector
9-pin subminiature D
baud rate
1200 to 19200
Protocol
8 data bits, 1 stop bit, no parity
Analog output
Number of channels
1
Connector
two-terminal screw clamp
Voltage range
–1.25 to +1.25V
Voltage resolution
0.0006V
Digital resolution
12 bits
Maximum source resistance
1
Ω
Maximum output current
2 mA
Linearity
±0.0006V
Typical accuracy
†
±2% ±0.02V
Isolation
grounded
Recommended operating temperature range
18 to 28°C (64 to 82°F)
Absolute operating temperature range
5 to 35°C (40 to 95°F)
Weight
2.2 lbs. (1 kg.)
†
Accuracy of the analog output can be significantly improved by calibration
Table 30
3560 Specifications
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...