6.5.2.3
INITiate:CONTinuous <Boolean>
This command starts continuous measurement if the <Boolean> parameter
value is ON or 1 or stops measurement if the <Boolean> parameter value is
OFF or 0. It is equivalent to selecting ON or OFF for the MEASURE parameter
with the
MEAS
soft-key function in the
INPUT
menu. The measurement status
indication on the display is updated to reflect any changes caused by the com-
mand. The 1560 is able to process new commands while measuring. The *RST
command sets the continuous measurement mode to OFF.
Example command: INIT:CONT ON
6.5.2.4
INITiate:CONTinuous?
This query command returns 1 if the measurement mode is continuous (MEA-
SURE: ON in the
MEAS
soft-key function) and returns 0 if the measurement
mode is OFF or COUNT.
Example command: INIT:CONT?
Example response: 1
6.5.2.5
SENSe:AVERage:COUNt <numeric_value>
This command sets the count for the moving average filter. It is equivalent to
setting COUNT for the
AVER
function in the
INPUT
soft-key menu. The
75
6 Digital Communications Interface
Commands
Command
Action
ABORt
Abort the measurement
INITiate[:IMMediate]
Start a series of COUNT N measurements
INITiate:CONTinuous <Boolean>
Set continuous measuring
INITiate:CONTinuous?
Query continuous measuring
SENSe[n]:AVERage:COUNt <numeric_value>
Set the count for input averaging
SENSe[n]:AVERage:COUNt?
Query the count for input averaging
SENSe[n]:AVERage[:STATe ] <Boolean>
Set averaging state ON or OFF
SENSe[n]:AVERage[:STATe ]?
Query averaging state
TRIGger[:SEQuence]:COUNt <numeric_value>
Set number of samples for COUNT N mode
TRIGger[:SEQuence]:COUNt?
Query number of samples for COUNT N mode
TRIGger[:SEQuence]:DELay <numeric_value>
Set measurement delay in seconds
TRIGger[:SEQuence]:DELay?
Query measurement delay
TRIGger[:SEQuence]:TIMer <numeric_value>
Set sequence timer value in seconds
TRIGger[:SEQuence]:TIMer?>
Query sequence timer value
Table 11
Measurement Control Commands
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...