The FORM FEED parameter can be used to eject a partially-printed page of pa-
per. Use the
LR
buttons to change the parameter to YES then press
ENTER
.
14.7.3
Printing Measurement Data
The 1560 can be programmed to print out measurement data in real time. The
printing of measurements is enabled using the
PRINT OUTPUT
function in
the
OUTPUT
soft-key menu (see Section 5.3.3). A window will show a list of
printer devices. Move to the PARP device using the
UD
buttons. Change the
option to ON using the
LR
buttons. The 1560 will immediately begin print-
ing all new measurements to the printer or terminal connected to the parallel
printer port of the Communication Module.
The 1560 can also print previously-acquired measurement data. This is done
using the
PRINT MEMORY
function in the
OUTPUT
soft-key menu (see
Section 5.3.4). A window will show a list of printer devices. Select the Parallel
Printer Device by moving to the PARP device using the
UD
buttons and
pressing
ENTER
. Now, enter the number of measurements to print using the
numeric buttons and pressing
ENTER
. Press
ENTER
again to confirm. The
number of measurements you specified, if available, will be retrieved from
memory and printed to the printer or terminal connected to the parallel printer
port of the Communication Module.
14.7.4
Device Setup Commands
The SYST:MODn:DEVn:READ? and SYST: MODn:DEVn:WRIT commands
(see Sections 6.5.8.19 and 6.5.8.20) can be used to read or set the setup param-
eters for the 3560 Parallel Printer Device. The device number of the Parallel
Printer Device is 4. The module number for the 3560 module is its position in
the stack. For example, if the 1560 has four add-on modules, the 3560 being
the last, the module number for the 3560 is 4. Table 34 shows the device com-
mands used to read or set the setup parameters of the Parallel Printer Device.
Below are some examples of using the device commands to set the device pa-
rameters. For these examples it is assumed that the 3560 Communication Mod-
ule is placed fourth in the stack. These commands can be sent through any
communication interface including the IEEE-488 and RS-232 interfaces.
SYST:MOD4:DEV4:WRIT “LFEE”,OFF
Disable linefeed.
1560 Thermometer Readout
User’s Guide
172
Device Command
Parameter
Description
LFEE
OFF or ON
linefeed
FFEE
YES
form feed
Table 34
Parallel Printer Device Commands
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...