CALIBRATE DEVICE: THRS
0 ADJ: 0.0
10K ADJ: 0.0
100K ADJ: 0.0
CAL DATE: 09-21-95
SER NUM: A32123
For the 2564 Thermistor Scanner Module the list of parameters includes
0_ADJ, 10K_ADJ, and 100K_ADJ as described above. The list also includes
the calibration date (CAL DATE) parameter, used to record the date the module
was calibrated, and the serial number parameter (SER NUM), used to record
the serial number of the module. New values can be entered for the parameters
using the numeric buttons and pressing
ENTER
. The
UD
buttons can be used
to move between parameters.
11.4.3
Calibration Procedure
Calibration requires adjustment of the 0_ADJ, 10K_ADJ, and 100_ADJ param-
eters at three specific input resistances. If the resistances used are approxi-
mately 0
Ω
, 10 k
Ω
and 100 k
Ω
respectively the adjustments are independent
and the procedure is simple. The order in which the adjustments are performed
is important. The adjustment of the 100K_ADJ parameter must be done last as
the adjustments of 0_ADJ and 10K_ADJ affect the measurement at 100 k
Ω
but
100K_ADJ does not affect the measurements at 0
Ω
or 10 k
Ω
. Any channel can
be used for calibration. Set the conversion type to R(
Ω
) to display resistance
(see Sec. 4.2.1.1). The accuracy required of the resistance standards is 1/4 of
the instrument accuracy; that is 0.05
Ω
at 0
Ω
, 0.25
Ω
(25 ppm) at 10 k
Ω
, and
2.5
Ω
(25 ppm) at 100 k
Ω
. The recommended procedure is as follows:
1.
Connect a 0
Ω
resistor to the input and measure its resistance. If a short-
ing wire is used the wire should run from pin 3 from the left to pin 4 to
pin 5 then back across to pin 2. Note the average error in the measure-
ment. Adjust the 0_ADJ parameter by subtracting the measured error.
For example, if the input is exactly 0.0
Ω
and readout shows –0.11
Ω
,
0_ADJ should be adjusted by adding 0.11 to it.
2.
Connect a 10 k
Ω
resistor (25 ppm accuracy) to the input and measure its
resistance. Note the average error in the measurement. Adjust the
10K_ADJ parameter by subtracting the measured error. For example, if
the input is exactly 10.00000 k
Ω
and the readout shows 10.00295 k
Ω
,
10K_ADJ should be adjusted by subtracting 2.95 from it.
1560 Thermometer Readout
User’s Guide
144
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...