4.
Connect the voltage source to Input 2, set it for 100 mV, and measure the
voltage with the 1560 (using channel 2) and the voltmeter simulta-
neously. Adjust the GA 2 parameter by subtracting the measured error.
5.
Verify the voltage accuracy of both input channels at 0 and 50 mV. The
accuracy must be within the short-term voltage accuracy given in the
specifications.
6.
Connect the calibrated thermocouple to Input 1 and insert the junction
into the temperature source. Measure the temperature with the 2565 (use
channel 1 and be sure to program the channel with the correct thermo-
couple type). Adjust the CJ 1 parameter by subtracting the measured er-
ror. For example, if the thermocouple temperature is actually 0.0°C but
the 1560 reads 0.184°C, the CJ OS 1 parameter should be adjusted by
subtracting 0.184 from it. If it was previously 0.000 it should now be
–0.184.
7.
Connect the calibrated thermocouple to Input 2. Measure the temperature
with the 2565 (use channel 2 and be sure to program the channel with
the correct thermocouple type). Adjust the CJ 2 parameter by subtracting
the measured error.
8.
Verify the temperature accuracy of both input channels using the cali-
brated thermocouple and temperature source. The accuracy should be
within 75% of the internal CJC accuracy given in the specifications.
9.
Record the date with the CAL DATE parameter.
1560 Thermometer Readout
User’s Guide
154
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...