Example response:0
6.5.5.3
OUTPut[n]:NCHannel <channel>
This command selects the input channel routed to the negative path of the out-
put channel. It is equivalent to setting -CHANNEL with the
OUTPUT CHAN
soft-key function (see Section 5.3.2). The output channel number is given by
the OUTP suffix
n
. If it is omitted output channel 1 is assumed. The input chan-
nel number is given by the <channel> parameter. The channel number is pre-
ceded by @ and enclosed in parentheses. For example, (@2) would be used to
select input channel 2. Use (@) to specify no input channel. This command
also sets the OUTPut:STATe to ON (see Section 6.5.5.7). The *RST command
sets the negative channel to none for all output channels.
Example command: OUTP1:NCH (@2)
6.5.5.4
OUTPut[n]:NCHannel?
This query command returns the input channel routed to the negative path of
the output channel. The output channel number is given by the OUTP suffix
n
.
If it is omitted output channel 1 is assumed. The returned channel number is
preceded by @ and enclosed in parentheses. If no channel is selected (@) is
returned.
Example command: OUTP1:NCH?
Example response: (@2)
6.5.5.5
OUTPut[n]:PCHannel <channel>
This command selects the input channel routed to the positive path of the out-
put channel. It is equivalent to sCHANNEL with the
OUTPUT CHAN
menu function (see Section 5.3.2). The output channel number is given by the
OUTP suffix
n
. If it is omitted output channel 1 is assumed. The input channel
number is given by the <channel> parameter. The channel number is preceded
by @ and enclosed in parentheses. For example, (@2) would be used to select
89
6 Digital Communications Interface
Commands
Number
Type
0
temperature
1
average
2
standard deviation
3
minimum
4
maximum
5
spread
6
n
7
input
Table 17
Output Calculation Types
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...