5.2.1.8
Thermistor R(T) Conversion
EDIT STHR 01
PROBE SER#: 1
CONVERSION: THRM-R(T)
B[0]: -35.540960
B[1]: 0.36568108
B[2]: -1.884784E-4
B[3]: 7.26691E-6
The following Steinhart-Hart equation is used for the thermistor R(T)
conversion:
r T K
B
B T
B T
B T
( [ ])
exp[
]
=
+
+
+
−
−
−
0
1
1
2
2
3
3
The user-defined parameters for the thermistor R(T) conversion are B[0] (
B
0
)
through B[3] (
B
3
). If
B
2
is not used, as is the case with some calibrations, B[2]
should be set to 0.0.
5.2.1.9
Thermocouple Volts Conversion
CHANNEL: TCS 3
PROBE SER#: 1
CONVERSION: V
The volts conversion for thermocouples displays the measurement in volts
rather than temperature.
The voltage readings are not compensated with the CJC value.
Note:
When selecting volts as the conversion type for thermocouples, no
cold-junction compensation is available. The value measured and displayed is
the uncompensated voltage as sensed at the input of the module.
1560 Thermometer Readout
User’s Guide
40
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...