4
General Operation
This section explains basic operation of the 1560
Black Stack
. Operation of the
1560 is explained in greater detail in subsequent sections: Section 5 explains
each of the functions available with the soft-keys and Section 6 explains the
communication commands used to operate the 1560 remotely.
4.1
Installing New Modules
The 1560 is generally supplied with certain modules already attached. How-
ever, new modules can be purchased later and attached to the back of the exist-
ing system to add extra channels or new capabilities. Modules are individually
calibrated at the factory before shipping and are ready to install and operate
without configuring or programming coefficients. They need only be attached
to the 1560.
Note:
Appropriate probe coefficients need to be entered.
NOTE:
Before proceeding to install a new module you must recognize
that adding new modules that contain input channels will alter the existing
arrangement of input channels. All probe characterizations stored in mem-
ory will be invalidated. After installing the new module you will have to
select the conversion algorithm and enter the characterization coefficients
again for each input channel you want to use. Be prepared by having the
coefficients on hand for each probe.
21
4 General Operation
Installing New Modules
Figure 6
Attaching new modules
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...