SET UP DEVICE: SERP
BAUD RATE: 2400
LR
LINEFEED: OFF
Use the
LR
buttons to change the baud rate, if necessary, and press
ENTER
. Setting the baud rate here also sets the baud rate of the Serial Com-
munication Device.
The LINEFEED parameter determines whether or not a linefeed character
(ASCII 0A hex or 10 decimal) is sent at the end of each line of transmission in
addition to a carriage return character (ASCII 0D hex or 13 decimal). Use the
LR
buttons to change the parameter then press
ENTER
.
14.6.2
Printing Measurement Data
The 1560 can be programmed to print out measurement data in real time. The
printing of measurements is enabled using the
PRINT OUTPUT
function in
the
OUTPUT
soft-key menu (see Section 5.3.3). A window will show a list of
printer devices. Move to the SERP device using the
UD
buttons. Change the
option to ON using the
LR
buttons. The 1560 will immediately begin print-
ing all new measurements to the printer or terminal connected to the RS-232
port of the Communication Module.
The 1560 can also print out previously-acquired measurement data. This is
done using the
PRINT MEMORY
function in the
OUTPUT
soft-key menu
(see Section 5.3.4). A window will show a list of printer devices. Select the Se-
rial Printer Device by moving to the SERP device using the
UD
buttons and
pressing
ENTER
. Now, enter the number of measurements to print using the
numeric buttons and pressing
ENTER
. Press
ENTER
again to confirm. The
number of measurements you specified, if available, will be retrieved from
memory and printed to the printer or terminal connected to the RS-232 port of
the Communication Module.
14.6.3
Device Setup Commands
The SYST:MODn:DEVn:READ? and SYST: MODn:DEVn:WRIT commands
(see Sections 6.5.8.19 and 6.5.8.20) can be used to read or set the setup param-
eters for the 3560 Serial Printer Device as explained in Section 14.5.4 above.
1560 Thermometer Readout
User’s Guide
170
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...