Unrecognized commands or commands with incorrect syntax or invalid param-
eters generate error messages in the error queue as explained in Section
6.5.10.22.
The 1560 does not allow compound commands (multiple commands per line
separated with semicolons).
All commands are sequential. The execution of each command is completed
before subsequent commands are processed.
6.4
Command Summary
An alphabetical listing of the commands implemented by the 1560 are shown
in Table 5 starting on page 66. All commands are available with the current
firmware version. Some commands may not be available with previous ver-
sions. (See Section 5.5.4 to determine the version number using the
MODULE
INFO
soft-key.) The section under which the command is explained is given
for each command.
6.5
Commands
This section explains each of the commands that can be used with the 1560.
The commands are arranged into the following groups:
•
Measurement Data Commands
•
Measurement Control Commands
•
Input Channel Commands
•
Probe Commands
•
Output Channel Commands
•
Printer Commands
•
Communication Interface Commands
•
Module Commands
•
System Commands
•
Status Commands
6.5.1
Measurement Data Commands
This group of commands deals with reading measurement data from the 1560.
These commands are summarized in Table 9.
6.5.1.1
CALCulate[n]:AVERage[n]:CLEar
This command clears the statistical functions of the specified channel. The
CALCulate suffix number specifies the channel number. Its range is 1 to 96. If
it is omitted it is assumed to be channel 1. The AVERage suffix number, if
65
6 Digital Communications Interface
Command Summary
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...