6.5.4.1
CALCulate[n]:CONVert:CATalog?
This query command returns a list of conversion types available for the given
input channel. The channel number is given by the CALC suffix
n
. The re-
sponse list gives the names of the conversion types as strings (enclosed in dou-
ble quotes) separated by commas. The available conversion types depend on the
type of module to which the channel belongs. Table 14 lists the names of con-
version types available with each input module type.
Example command: CAL2:CONV:CAT?
Example response:
“I90”, ”RES”,”W”,”I68”,”CVD”,”POLY”
Input Type
Conversion types
Mnemonic
PRT/SPRT
ITS-90 (default)
I90
R(
Ω
)
RES
W(T90)
W
IPTS-68
I68
CVD
CVD
POLYNOMIAL
POLY
Thermistor
THRM-R(T) (default)
TRES
R(
Ω
)
RES
THRM-T(R)
TTEM
POLYNOMIAL
POLY
2564 module
ITS-90
I90
R(
Ω
)
RES
W(T90)
W
ITPS-68
I68
CVD
CVD
POLYNOMIAL
POLY
THRM-R(T)
TRES
THRM-T(R)
TTEM
Thermocouple
TC-K (default)
K
VOLTS
VOLT
TC-B
B
TC-E
E
TC-J
J
1560 Thermometer Readout
User’s Guide
82
Table 14
Conversion Type Mnemonics
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...