5322A
Calibration Manual
30
5. Select the first position with a nominal value of 10 k
Ω
.
6. Measure its value with a standard meter.
7. Write down the new value. Use the cursor keys and the numerical keypad to enter the new
value.
8. Confirm the value with the
Write
softkey. If there is no need to change the value, push the
Exit
softkey to keep the original value.
9. Adjust each of the constants in the table.
10. When finished with the adjustment, push the
Exit
softkey repeatedly until the basic display
with the Low resistance source function shows.
This completes this part of the adjustment.