Multifunction Electrical Tester Calibrator
List of Error Messages
89
ID
Error message
Description
653
Compensator
overload
Overloaded compensator (high residual impedance,
measurement current or long measurement time).
654
Compensator
disabled
Cannot select compensator because residual
impedance is >10
Ω
.
655
Cont. current too
high
Continuous current is too high.
656
Short-t. current too
high
Short-term current is too high.
661
Test voltage too
high.
External test voltage is too high. Use lower test voltage
on DUT.
662
Test voltage
unstable.
Unstable external test voltage.
701
Output/Input
overloaded
Input or output signal is over specified limits. Decrease
signal level.
701
Output/Input
overloaded
Input or output signal over specified limits. Decrease
signal level.
703
Temperature too
high.
Product power stage overheated. Disconnect external
load.
704
Output/Input
overloaded
Input or output signal over specified limits. Decrease
signal level.
705
Output/Input
overloaded
Input or output signal over specified limits. Decrease
signal level.
706
Output/Input
overloaded
Input or output signal over specified limits. Decrease
signal level.
711
Value too large.
Set up value is too large. Set value in specified limits.
712
Value too small.
Set up value is too low. Set value in specified limits.
713
Negative value.
Negative value is not permitted. Do not set a negative
value.
715
Rescan is not ready Temperature overheating of scan circuit.
716
Rescan required
Required impedance correction mode cannot be
selected without Rescan. Do Rescan first.
721
Unknown function. Error in internal communication.
722
Unexpected
crossing.
Error in internal communication.
731
Calibrator not
ready
Error in internal communication.
732
Internal cpu RESET Product will be restarted.
Table 34. Error Messages (cont.)