Multifunction Electrical Tester Calibrator
Adjustment Procedures
25
9. Confirm the value with the
Write
softkey. If there is no need to change value, push the
Exit
softkey to keep the original value.
10. Go through all positions in the list and make an adjustment for each resistor.
11. When you finish adjustment, push the
Exit
softkey repeatedly until the basic display with
the Low resistance source function shows.
This completes this part of the adjustment.
Table 4. Low Resistance Source Adjustment Points
Position
Adjustment
Point
Nominal
Calibration
Value
Required
Accuracy of
Adjustment
Lower Limit
Upper Limit
LVR00
0.01
Ω
0.01
Ω
0.3 %
-30 %
+30 %
LVR01
0.14
Ω
0.1
Ω
1.0 %
-50 %
+50 %
LVR02
0.24
Ω
0.2
Ω
0.2 %
-20 %
+20 %
LVR03
0.43
Ω
0.4
Ω
0.1 %
-10 %
+10 %
LVR04
0.77
Ω
0.8
Ω
0.02 %
-10 %
+10 %
LVR05
1
Ω
1.0
Ω
0.02 %
-10 %
+10 %
LVR06
1.9
Ω
2.0
Ω
0.02 %
-10 %
+10 %
LVR07
3.6
Ω
3.4
Ω
0.02 %
-10 %
+10 %
LVR08
5.4
Ω
6.0
Ω
0.02 %
-10 %
+10 %
LVR09
10
Ω
10
Ω
0.02 %
-10 %
+10 %
LVR10
18
Ω
20
Ω
0.02 %
-10 %
+10 %
LVR11
36
Ω
33
Ω
0.02 %
-10 %
+10 %
LVR12
55
Ω
55
Ω
0.02 %
-10 %
+10 %
LVR13
100
Ω
100
Ω
0.02 %
-10 %
+10 %
LVR14
190
Ω
200
Ω
0.02 %
-10 %
+10 %
LVR15
360
Ω
400
Ω
0.02 %
-10 %
+10 %
LVR16
600
Ω
600
Ω
0.02 %
-10 %
+10 %
LVR17
1 k
Ω
1 000
Ω
0.02 %
-10 %
+10 %
LVR18
1.9 k
Ω
2 000
Ω
0.02 %
-10 %
+10 %
LVR19
3.5 k
Ω
3 500
Ω
0.02 %
-10 %
+10 %
LVR20
7.1 k
Ω
6 800
Ω
0.02 %
-10 %
+10 %
LVR21
11 k
Ω
11 000
Ω
0.02 %
-10 %
+10 %
LVR22
20 k
Ω
20 000
Ω
0.05 %
-10 %
+10 %
LVR23
40 k
Ω
40 000
Ω
0.05 %
-10 %
+10 %