Multifunction Electrical Tester Calibrator
Adjustment Procedures
53
RCD Trip Current Adjustment
The RCD adjustment procedure includes adjustment of the internal ammeter and line voltage
meter. The ammeter has three ranges (30 mA, 300 mA, and 3000 mA) for RCD function, and
two ranges (30 mA and 150 mA) for RCD in PAT function. The line voltage meter is one range.
To adjust the RCD trip current, see
1. Connect the 5522A to the Product.
Figure 36. RCD Trip Current Adjustment Connections
2. Push the
Setup
softkey on the Product.
3. Highlight
Calibration
in the menu and push the
Select
softkey.
4. From the calibration menu, highlight
RCD Trip Current
and push the
Select
softkey. The
list of ranges shows, see
5. Highlight
RCD I01 Trip Current 30 mA
and push the
Select
softkey.
1
4
7
/
+
STBY
5522A CALIBRATOR
Fluke 5322A Fluke 5522A