5322A
Calibration Manual
84
Table 29. DC/AC Voltage Multimeter Test
Nominal Output
Voltage
Frequency
(Hz)
Lower Limit
(V)
Upper Limit
(V)
4 V
-
3.989
4.011
-4 V
-
-4.011
-3.989
9 V
-
8.982
9.019
40 V
-
39.870
40.130
-40 V
-
-40.130
-39.870
90 V
-
89.770
90.230
200 V
-
199.050
200.950
-200 V
-
-200.950
-199.050
400 V
-
398.650
401.350
600 V
-
598.250
601.750
800 V
-
797.850
802.150
1000 V
-
997.450
1002.550
9 V
50
8.982
9.019
9 V
60
8.982
9.019
9 V
400
8.982
9.019
90 V
50
89.770
90.230
90 V
60
89.770
90.230
90 V
400
89.770
90.230
200 V
50
199.050
200.950
200 V
60
199.050
200.950
200 V
400
199.050
200.950
400 V
60
398.650
401.350
600 V
60
598.250
601.750
800 V
60
797.850
802.150
1000 V
60
997.450
1002.550