Multifunction Electrical Tester Calibrator
Calibration
15
For example, if you select
Ground Bond Resistance
from the calibration menu, the Ground
Bond Resistance screen shows the screen in
. The screen shows
GBR R00
highlighted. To choose this parameter, push the
Select
softkey.
Figure 13. Ground Bond Resistance Screen
Types of Adjustment Points
A calibration procedure contains these types of adjustment points:
1. Points where an external standard measures the internal resistance. The standard unit then
manually writes and saves this value to the Product. These functions use this system:
• Ground Bond Resistance
•
Low Resistance Decade
• High Resistance Decade
Содержание 5322A