Multifunction Electrical Tester Calibrator
Adjustment Procedures
23
Table 3. 4-wire (4W) Adjustment Points
The maximum acceptable deviation of the calibration value to the nominal value of each
resistor is in the Lower Limit and Upper Limit columns. If any of the partial resistors is out of
shown limits, it must be replaced. Replacement should be done only by an authorized service
center.
Nominal
Resistance
Value
Position
Required
Accuracy of
Adjustment
Lower Limit
Upper Limit
170
Ω
R12
GBR2W
±
200 m
Ω
-10 %
+10 %
470
Ω
R13
GBR2W
±
500 m
Ω
-10 %
+10 %
900
Ω
R14
GBR2W
±
1
Ω
-10 %
+10 %
1700
Ω
R15
GBR2W
±
2
Ω
-10 %
+10 %
Nominal
Resistance
Value
Position
Required
Accuracy of
Adjustment
Lower Limit
Upper Limit
15 m
Ω
R00
GBR4W
±
0.1 m
Ω
-50 %
+50 %
45 m
Ω
R01
GBR4W
±
0.2 m
Ω
-50 %
+50 %
90 m
Ω
R02
GBR4W
±
0.4 m
Ω
-30 %
+30 %
330 m
Ω
R03
GBR4W
±
0.5 m
Ω
-20 %
+20 %
500 m
Ω
R04
GBR4W
±
0.8 m
Ω
-10 %
+10 %
900 m
Ω
R05
GBR4W
±
1 m
Ω
-10 %
+10 %
1.7
Ω
R06
GBR4W
±
2 m
Ω
-10 %
+10 %
4.7
Ω
R07
GBR4W
±
5 m
Ω
-10 %
+10 %
9
Ω
R08
GBR4W
±
15 m
Ω
-10 %
+10 %
17
Ω
R09
GBR4W
±
30 m
Ω
-10 %
+10 %
47
Ω
R10
GBR4W
±
70 m
Ω
-10 %
+10 %
90
Ω
R11
GBR4W
±
100 m
Ω
-10 %
+10 %
170
Ω
R12
GBR4W
±
200 m
Ω
-10 %
+10 %
470
Ω
R13
GBR4W
±
500 m
Ω
-10 %
+10 %
900
Ω
R14
GBR4W
±
1
Ω
-10 %
+10 %
1700
Ω
R15
GBR4W
±
2
Ω
-10 %
+10 %
1 m
Ω
R16
GBR4W
±
0.05 m
Ω
-20%
+20 %
Table 2. 2-Wire (2W) Adjustment Points (cont.)