Multifunction Electrical Tester Calibrator
Adjustment Procedures
29
Once the process starts, adjust all partial resistors to keep the specified accuracy.
The maximum acceptable deviation of the calibration value to the nominal value of each
resistor is in the Lower Limit and Upper Limit columns. If any of the partial resistors is out of the
shown limits, it must be replaced. Replacement should be done only by an authorized service
center.
High Resistance Source Adjustment 1.5 kV
Fully adjust the low resistance source with the front panel keyboard. Calibration is based on dc
measurements of resistance in the test points in
. Part of the adjustment sets the test
voltage voltmeter and the adjustment of an external option, Resistance multiplier, respectively.
Use the 8588A and 1653 or similar for these procedures, see
.
1. Connect the Product to the 1653 (for values >10 M
Ω
) or to 8588A (for values
≤
10 M
Ω
), see
.
2. Push the
Setup
softkey on the Product.
3. Select
Calibration
and confirm with the
Select
softkey.
4. From the calibration menu, highlight
High Resistance Source
and push
E
.
The list of adjustment points shows in this format:
HVR Rxx nominal resistance value
where HVR is a symbol of the function,
Rxx is the ordinal number of the point
20.0 M
Ω
HVR12
0.1 %
-10 %
+10 %
40.0 M
Ω
HVR13
0.1 %
-10 %
+10 %
78.0 M
Ω
HVR14
0.1 %
-10 %
+10 %
150 M
Ω
HVR15
0.1 %
-10 %
+10 %
287 M
Ω
HVR16
0.2 %
-10 %
+10 %
530 M
Ω
HVR17
0.2 %
-10 %
+10 %
900 M
Ω
HVR18
0.2 %
-10 %
+10 %
1.4 G
Ω
HVR19
0.2 %
-10 %
+10 %
2.9 G
Ω
HVR20
0.5 %
-10 %
+10 %
5.4 G
Ω
HVR21
0.5 %
-10 %
+10 %
10 G
Ω
HVR22
0.5 %
-10 %
+10 %
18 G
Ω
HVR23
0.5 %
-10 %
+10 %
33 G
Ω
HVR24
1.0 %
-10 %
+10 %
50 G
Ω
HVR25
1.0 %
-10 %
+10 %
Table 5. High Resistance Source Adjustment Points (cont.)