5322A
Calibration Manual
22
2. Push the Product
Setup
softkey.
3. Select
Calibration
and push the
Select
softkey.
4. From the calibration menu, select
Ground Bond Resistance 2W
or
Ground Bond
Resistance 4W
and push the
Select
softkey.
5. The list of calibration points shows in this format:
GBR2W(4W)
Rxx nominal resistance value
where
GBR is symbol of the function
Rxx is ordinal number of the point
6. Select and measure the first resistance position with a standard multimeter.
7. Write down the new adjustment value. Use the cursor keys and the numerical keypad to
enter the new adjustment value.
8. Confirm the value with the
Write
softkey. To not change the adjustment value, push the
Exit
softkey to keep the original value.
9. Go through all steps (constants) in
and make a new adjustment for each of the
resistors.
10. When finished with adjustment, push the
Exit
softkey repeatedly until the basic display
shows.
This completes this part of the adjustment.
Table 2. 2-Wire (2W) Adjustment Points
Nominal
Resistance
Value
Position
Required
Accuracy of
Adjustment
Lower Limit
Upper Limit
25 m
Ω
R00
GBR2W
±
3 m
Ω
-50 %
+50 %
55 m
Ω
R01
GBR2W
±
3 m
Ω
-50 %
+50 %
90 m
Ω
R02
GBR2W
±
3 m
Ω
-30 %
+30 %
330 m
Ω
R03
GBR2W
±
5 m
Ω
-20 %
+20 %
500 m
Ω
R04
GBR2W
±
5 m
Ω
-10 %
+10 %
900 m
Ω
R05
GBR2W
±
5 m
Ω
-10 %
+10 %
1.7
Ω
R06
GBR2W
±
7 m
Ω
-10 %
+10 %
4.7
Ω
R07
GBR2W
±
10 m
Ω
-10 %
+10 %
9
Ω
R08
GBR2W
±
20 m
Ω
-10 %
+10 %
17
Ω
R09
GBR2W
±
30 m
Ω
-10 %
+10 %
47
Ω
R10
GBR2W
±
70 m
Ω
-10 %
+10 %
90
Ω
R11
GBR2W
±
100 m
Ω
-10 %
+10 %