5322A
Calibration Manual
14
• RCD Trip Current
Calibration data of ammeter and line voltage meter in the RCD function.
• Leakage Current
Calibration data of internal leakage current ammeter.
• HV Probes
Calibration constants for 10 kV and 40 kV external probes.
• High Resistance Multiplier
Calibration constants external option high resistance multiplier.
Access the Calibration Procedure
Use the calibration code to access the Product calibration procedure. The default calibration
password is
235
. In
General Setup
, if the calibration password is
0
, access to the calibration
procedure unprotected with the password.
To enter the password:
1. Push the
Setup
softkey to open the setup menu.
2. Select
Calibration
and confirm with the
Select
softkey. The Product requests the
calibration code.
3. Enter the calibration code with the numerical keypad and push
E
. If you enter an
incorrect calibration code, an error message shows:
Bad calibration code!
4. When you enter the correct calibration code, the Product shows the calibration menu (see
Select the Calibration Type
To select the calibration type:
From the calibration menu, use the cursor keys or the rotary knob to move the cursor through
the listed items shown in
.