5322A
Calibration Manual
20
For example,
shows a 10 kV HV Probes screen.
Figure 18. 10 kV Probe Screen
Identification of the active resistance segment shows in this syntax:
10kV probe ratio
Dividing constant K for 10 kV probe
1000.0
Adjustment value of dividing ratio in terms of 1/K
At the bottom of the screen are two softkeys:
Write
New value added into memory, the old value is
irreversibly lost.
Exit
Current calibration process terminates. When pushed,
the calibration memory holds the last value (old or new
value entered) and the Product returns to the
calibration menu. It is not necessary to adjust all
ranges. Adjustment of only selected ranges is
possible.
Terminate the Calibration Process
You can terminate the calibration process under any of these instances:
•
Complete or partial adjustment (for example, specific steps in the calibration procedure) is
complete after new calibration data was entered.
• Calibration is started but no calibration data is entered.
Push the
Exit
softkey to go to the previous level of the calibration menu.
Note
Some calibration functions require that the output terminals be switched on manually.
Push
O
to activate the output terminals. Description of this repeated step is omitted
in the procedures listed in this manual.