Multifunction Electrical Tester Calibrator
Adjustment Procedures
51
AC Range 5000 V Adjustment
To make this adjustment, see
1. Connect the high voltage source to the HV and COM terminals of the Product.
2. Connect the high voltage probe input parallel to the high voltage source output terminals.
3. Set the multimeter to the dc voltage function and connect the it to the high voltage probe
output.
4. Select
Range 5000V 55 Hz low (100V)
from the calibration menu and push the
Select
softkey.
5. Set the output voltage to 100.0 V dc on the high voltage source and switch the output
terminals ON.
6. Use the rotary knob or cursor keys to set the value the Product to show 100.0 V.
7. Push the
Write
softkey to confirm the new value.
8. Select
Range 5000V high (+5000 V)
and push the
Select
softkey.
9. Set the output voltage to 5000 V dc on the high voltage source and switch the output
terminals ON. The Product shows the approximate nominal value.
10. Push the
Write
softkey to save the new value.
This completes this part of the adjustment.