9 Measurement
FLUXUS G722ST-HT
9.3 Start of the measurement
97
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
Input of the max. flow
The expected max. flow is needed to calculate the recommended distance between the measuring barriers.
• Enter the expected max. flow.
• Press ENTER.
Adjustment of the transducer distance for transducer pair A
The recommended transducer distance for transducer pair A will be displayed. The transducer distance is measured
between the inner edges of the transducers. On very small pipes, a negative transducer distance is possible.
• Mount the transducers of transducer pair A on the pipe and adjust the transducer distance.
• Press ENTER.
Adjustment of the distance between the measuring barriers
The recommended measuring barrier distance (distance between the transducers of the transducer pairs A and B) will be
displayed.
• Mount the transducers for transducer pair B on the pipe respecting the displayed distance between the measuring
barriers. The distance has to be adjusted with an accuracy of 1 mm. In case the recommended distance between the
measuring barriers displayed by the transmitter cannot be set because of the transducer size, mount the transducer
pair B at the smallest possible distance to transducer pair A. Make sure to keep a few millimeters distance between the
transducers of measuring barrier A and B. The transducers should not touch.
• Enter the distance between the measuring barriers.
• Press ENTER.
Check of the transducer distance for transducer pair B
The recommended transducer distance for transducer pair B will be displayed.
• Check whether the transducer distance of transducer pair B corresponds to the value displayed by the transmitter.
The transducer distances of transducer pair A and B have to be equal.
• Correct the transducer distance, if necessary.
• Press ENTER.
Parameter search
Before starting the measurement, the transmitter performs a parameter search. This process takes about 10 minutes.
The remaining time is displayed.
After completion of the parameter search, the measurement is started. The measured values are displayed.
Measurement\...\Expected max. flow
Measurement\...\Transducer distance
Notice!
The accuracy of the recommended transducer distance depends on the accuracy of the entered pipe and fluid
parameters.
Measurement\...\Barrier distance
Measurement\...\Transducer distance
Fig. 9.1:
Parameter search
Содержание FLUXUS G722ST-HT
Страница 1: ...FLUXUS G722ST HT UMFLUXUS_G722ST HTV1 2EN Operating instruction...
Страница 6: ...Table of contents FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 6...
Страница 173: ...Annex FLUXUS G722ST HT D Conformity declarations 173 UMFLUXUS_G722ST HTV1 2EN 2021 09 01 D Conformity declarations...
Страница 174: ...Annex D Conformity declarations FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 174...
Страница 175: ......