11 Maintenance and cleaning
11.1 Maintenance
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
106
11.1
Maintenance
The transmitter and the transducers are practically maintenance-free. In order to ensure security, the following
maintenance intervals are recommended:
11.2
Cleaning
Stainless steel housing
• Clean the housing with a soft cloth and care and cleaning spray for stainless steel.
Aluminum housing
• Clean the housing with a soft cloth. Do not use detergents.
Transducers
• Remove traces of coupling compound from the transducers with a soft paper towel.
11.3
Calibration
If installed as recommended in an appropriate location, used cautiously and serviced conscientiously, no troubles should
appear.
The transmitter has been calibrated at factory and, usually, a recalibration is not necessary.
A recalibration is recommended if:
• the contact surfaces of the transducers show visible wear
• the transducers were used for a prolonged period at high temperatures (several months > 200 °C for high temperature
transducers)
In order to realize a recalibration under reference conditions, either the transmitter, the transducers or both have to be sent
to FLEXIM.
11.4
Firmware update
A firmware update is carried out with FluxDiagReader or FluxDiag (option).
• For the connection of the transmitter to a PC, see section 7.6.
• For the operation of FluxDiagReader or FluxDiag see the help functions of these programs.
Tab. 11.1: Recommended maintenance intervals
item
maintenance step
interval
measure
stainless steel housing
• transmitter
• junction box
• transducer mounting fixture
visual inspection for
corrosion and damages
annually or more
frequently, depending on
the ambient conditions
cleaning
visual inspection for
contamination
annually or more
frequently, depending on
the ambient conditions
aluminum housing
• transmitter
visual inspection for
contamination
annually or more
frequently, depending on
the ambient conditions
transducers
check of the transducer
coupling on the pipe
annually
replacement of coupling foil,
if necessary
transmitter
check for firmware updates
annually
update, if necessary
transmitter
functional test
annually
reading of measured and
diagnostic values
transmitter and transducers
calibration
-
see section 11.3
Содержание FLUXUS G722ST-HT
Страница 1: ...FLUXUS G722ST HT UMFLUXUS_G722ST HTV1 2EN Operating instruction...
Страница 6: ...Table of contents FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 6...
Страница 173: ...Annex FLUXUS G722ST HT D Conformity declarations 173 UMFLUXUS_G722ST HTV1 2EN 2021 09 01 D Conformity declarations...
Страница 174: ...Annex D Conformity declarations FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 174...
Страница 175: ......