18 Advanced functions
FLUXUS G722ST-HT
18.6 Event protocol
147
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
18.5.4
Status display of the event triggers
The state of the event triggers is displayed during the measurement.
• Press
to scroll through the second line from below until the event trigger status is displayed.
The status display of the event triggers is structured as follows.
Rx =
with
X
being the number of the event trigger and
a pictogram according to Tab. 18.4.
18.6
Event protocol
If an error occurs, an error message will be displayed in the first line indicating the symbol
. The error message can be
displayed via the menu item
Event log
.
• Select the menu item
Event log
.
• Press ENTER.
A list is displayed containing all error messages since the last time the transmitter was switched on.
• Press
and
to select an error message.
• Press ENTER.
The display indicates the cause of the error.
Notice!
There is no visual or acoustic indication of event trigger switching.
Tab. 18.4: Pictograms for the status display of the event triggers
No.
Function
(switching condition)
Type
(holding behavior)
current state
R
=
1
MAX (x>limit)
Non-hold
closed
2
MIN (x<limit)
Hold
open
3
Within range
Hold for a while
4
Out of range
ERR (x=fail)
Example
R1 =
Special functions\System settings\Event log
Notice!
After reading out the event protocol the error message symbol will be deleted on the display, even if the error has not
been eliminated yet.
The event protocol will be deleted after a restart of the transmitter.
Содержание FLUXUS G722ST-HT
Страница 1: ...FLUXUS G722ST HT UMFLUXUS_G722ST HTV1 2EN Operating instruction...
Страница 6: ...Table of contents FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 6...
Страница 173: ...Annex FLUXUS G722ST HT D Conformity declarations 173 UMFLUXUS_G722ST HTV1 2EN 2021 09 01 D Conformity declarations...
Страница 174: ...Annex D Conformity declarations FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 174...
Страница 175: ......