16 Data logger
16.1 Configuration of the data logger
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
130
Storage rate
The storage rate is the frequency to transmit or store measured values. If a start time for storing the measured values is
defined, a storage rate has to be entered.
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Storage rate
is displayed.
• Select a storage rate from the scroll list.
• Press ENTER.
• If
Customized
is selected, enter the storage rate.
• Press ENTER.
Ringbuffer
The data logger can be configured as linear logger or ringbuffer. If the ringbuffer is deactivated and the data logger is full,
the storing of measured values will be terminated. It can be continued after clearing the data logger. If the ringbuffer is
activated and the data logger is full, the oldest measured values will be overwritten. In ringbuffer mode, the remaining
capacity of the data logger is displayed during the measurement, e.g.:
Log←→ : 1d 6h 57m
is displayed, if no measured values have been overwritten.
Log|←→| : 1d 6h 57m
is displayed, if the old measured values have been overwritten.
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Ringbuffer
is displayed.
• Select
On
to activate the ringbuffer.
• Press ENTER.
If the ringbuffer is deactivated and the data logger is full, the storing of measured values will be terminated.
• Select
Off
to deactivate the ringbuffer.
• Press ENTER.
Storage mode
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Storage mode
is displayed.
• Press ENTER.
• Select
Sample
to store the current measured value.
• Select
Average
if the average of all undamped measured values of a storage interval is to be stored.
Special functions\Data logger\Configuration\Storage rate
Special functions\Data logger\Configuration\Ringbuffer
Special functions\Data logger\Configuration\Storage mode
Notice!
The storage mode does not affect the outputs.
Notice!
Storage mode
=
Average
The average of the physical quantity and of other quantities assigned to the measuring channel, e.g., the measured
temperature, will be calculated.
If a storage rate < 5 s is selected,
Sample
will be used.
If no average could be calculated over the complete storage interval, the value will be marked as invalid.
Содержание FLUXUS G722ST-HT
Страница 1: ...FLUXUS G722ST HT UMFLUXUS_G722ST HTV1 2EN Operating instruction...
Страница 6: ...Table of contents FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 6...
Страница 173: ...Annex FLUXUS G722ST HT D Conformity declarations 173 UMFLUXUS_G722ST HTV1 2EN 2021 09 01 D Conformity declarations...
Страница 174: ...Annex D Conformity declarations FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 174...
Страница 175: ......