6 Mounting
6.2 Transducers
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
36
Step 4: Fixation of the rail to the pipe
• Place the tension strap clamp (2) in the rail (see Fig. 6.21). Observe the orientation of the tension strap clamp.
• Slightly tighten the nut of the tension strap clamp (2).
• Screw the rail to tension strap clamp (1) (see Fig. 6.22).
• Tighten the nut of tension strap clamp (1), but not too firmly in order not to damage the tension strap.
• Repeat the steps to fix the second rail on the opposite side of the pipe. Make sure that the 2 tension strap clamps are as
exactly as possible opposite each other. To do this, check whether the distance between the tension strap clamps is half
the pipe circumference.
• Select the installation instruction of the supplied clasp:
Fig. 6.20: Ratchet clasp with tension strap
1 – sense of rotation
2 – edge
3 – lever
4 – clasp screw with slot
Fig. 6.21: Rail with tension strap clamp
1 – nut
2 – tension strap clamp
Fig. 6.22: Rail, fixed on one side to the pipe
1 – tension strap clamp
2 – tension strap clamp
3 – nut
1
2
4
3
4
1
2
2
1
3
1
Содержание FLUXUS G722ST-HT
Страница 1: ...FLUXUS G722ST HT UMFLUXUS_G722ST HTV1 2EN Operating instruction...
Страница 6: ...Table of contents FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 6...
Страница 173: ...Annex FLUXUS G722ST HT D Conformity declarations 173 UMFLUXUS_G722ST HTV1 2EN 2021 09 01 D Conformity declarations...
Страница 174: ...Annex D Conformity declarations FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 174...
Страница 175: ......