![Flexim FLUXUS G722ST-HT Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/flexim/fluxus-g722st-ht/fluxus-g722st-ht_operating-instruction_2304717030.webp)
6 Mounting
6.2 Transducers
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
30
6.2.2
Installation of the transducers
6.2.2.1
Orientation of the transducers and determination of the distance between the transducers and
measuring barriers
Observe the orientation of the transducers. If the transducers of a transducer pair have been mounted properly, the
engravings on them form an arrow pointing in the flow direction. The transducer cables show in opposite directions.
The transducer distance is measured between the inner edges of the transducers of a measuring barrier.
The measuring barrier distance is measured between the inner edges of the emitters and/or receivers of measuring
barrier A and B.
• Select the installation instructions of the supplied transducer mounting fixture.
Fig. 6.8:
Orientation and distance of the transducers
Δx – measuring barrier distance
a – transducer distance
A
1
B
1
A
2
B
2
Δx
a
a