Reset of measured values
• Select the list item
Reset meas. val.
• Press ENTER.
The measured value output simulates a reposing application for the duration of the signal. The actual measured flow
velocity is ignored and the measured value is set to zero. All values of the physical quantity derived from the flow velocity
and the values from the calculation channels also yield zero.
The transmitter continues the measurement if the condition of the remote function is no longer met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
Reset of the totalizers
• Select the list item
Reset totalizers
.
• Press ENTER.
The totalizers are set to zero.
Totalizing starts at zero again, as soon as the condition for the remote function is no longer met.
When the totalizers are reset to zero with the remote function, the character
H
is displayed next to the measured value
during the measurement.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
Stop of the totalizers
• Select the list item
Stop totalizers
.
• Press ENTER.
The totalizers are stopped for the duration of the signal.
Totalizing will be continued with the last registered totalizer value if the condition of the remote function is no longer met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
Содержание FLUXUS G722ST-HT
Страница 1: ...FLUXUS G722ST HT UMFLUXUS_G722ST HTV1 2EN Operating instruction...
Страница 6: ...Table of contents FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 6...
Страница 173: ...Annex FLUXUS G722ST HT D Conformity declarations 173 UMFLUXUS_G722ST HTV1 2EN 2021 09 01 D Conformity declarations...
Страница 174: ...Annex D Conformity declarations FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 174...
Страница 175: ......