13 User modes
FLUXUS G722ST-HT
13.3 SuperUser mode and SuperUser ext. mode
115
UMFLUXUS_G722ST-HTV1-2EN, 2021-09-01
Example
factor: 1.1
offset: -10 cm/s = -0.1 m/s
If a flow velocity v = 5 m/s is measured, before the calculation of the derived quantities it will be corrected as follows:
v
cor
= 1.1 · 5 m/s - 0.1 m/s = 5.4 m/s
Example
factor: -1
offset: 0
Only the sign of the measured values changes.
Notice!
The correction data will not be stored until a measurement is started.
If the transmitter is switched off without starting a measurement, the entered correction data will be lost.