9 Measurement
9.2 Measurement settings
FLUXUS G722ST-HT
2021-09-01, UMFLUXUS_G722ST-HTV1-2EN
88
9.2.6
Configuration of an output
If the transmitter is equipped with outputs, they have to be
configured. The measured value, the status value or an
event value can be transmitted via the different outputs.
In the following, the configuration of an analog output is
described.
• Select the program branch
Options
.
• Press ENTER.
Assignment of an output
• Select the menu item
Outputs
.
• Press ENTER.
• Select the output (here:
Current I1(--)
).
• Press ENTER.
The scroll list contains all available outputs of the transmitter:
– Current Ix (--)
– Voltage Ux (--)
– Digital output
Bx (--)
If the output has already been enabled, it is displayed as follows:
I1 (
)
.
• Select
Yes
to change the settings for an already enabled output.
• Select
No
to cancel the assignment and to return to the previous menu item.
• Press ENTER.
Assignment of a source item
A source item has to be assigned to each selected output.
• Select a source item whose measured value, status value or event value is to be transmitted to the output.
• Press ENTER.
Options
Outputs
Select an output
Enable the output
Source item
see annex A
Options\Outputs\Current I1(--)
Options\Outputs\Current I1\Enable I1
Options\Outputs\...\Source item
Содержание FLUXUS G722ST-HT
Страница 1: ...FLUXUS G722ST HT UMFLUXUS_G722ST HTV1 2EN Operating instruction...
Страница 6: ...Table of contents FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 6...
Страница 173: ...Annex FLUXUS G722ST HT D Conformity declarations 173 UMFLUXUS_G722ST HTV1 2EN 2021 09 01 D Conformity declarations...
Страница 174: ...Annex D Conformity declarations FLUXUS G722ST HT 2021 09 01 UMFLUXUS_G722ST HTV1 2EN 174...
Страница 175: ......