Page 74
Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
Key sequ. /
Action
Detail of the display Explanation
Base
Make about 5 measurements at different lo-
cations of the base.
The mean value of all readings obtained for
the normalization will be displayed.
[Base material (Fe)]
: The measurements
should be made on base.
[Delete: DEL]
: Use
DEL
to delete the last
measurement
DEL
,
2x
DEL
to delete all
readings obtained for the normalization.
[OK: ENTER]
: Use
ENTER
to end and
store the normalization. The previous nor-
malization will be overwritten.
ENTER
[Entry:
]
: Use the arrow keys to set the
rated
-
ferrite content value of the calibra-
tion standard.
The rated value can be set faster if a mea-
surement is performed on the calibration
standard and then the rated value is correct-
ed using the arrow keys.
[CAL-rat. 1: 680]
: Display of the set rated
value for the
-
ferrite content of the calibra-
tion standard (Example: 0.680 Fe%)
[Cancel: ENTER]
: Using ENTER cancels
the calibration procedure
Pressing
FINAL-RES
:
Determination of the Normalized Coun-
trate Xn of a Calibration Standard During a
Master Calibration
Pressing
BLOCK-RES
:
Displaying Xn Ranges for Calibration Stan-
dards for the Master Calibration,
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...