Operator‘s Manual FERITSCOPE
®
FMP30
Page 65
Normalization, Calibration and Master Calibration
6.3.2 Corrective Calibration Procedure
Key sequ. /
Action
Detail of the display Explanation
CAL
Use
CAL
to start the corrective calibration
of the open Application.
CAL
appears and remains on the display
while the corrective calibration is per-
formed.
[s]
: Standard deviation
[n]
: Number of measurements
[Base material (Fe)]
: The measurements
should be made on base.
[Delete cal: DEL]
: Using
DEL
deletes the
corrective calibration (
[Skip: ENTER]
: Using
ENTER
skips the
normalization (the stored normalization is
retained unchanged).
[Cancel: CAL]
: Using
CAL
cancels the cor-
rective calibration (the stored corrective cal-
ibration is retained unchanged).
Base
Make about 5 measurements at different lo-
cations of the base.
[Delete: DEL]
: Use
DEL
to delete the last
measurement
,
2x
DEL
to delete all read-
ings obtained for the normalization.
[Cancel: CAL]
: Using
CAL
cancels the cor-
rective calibration (the stored corrective cal-
ibration is retained unchanged).
[OK: ENTER]
: Use
ENTER
to end and
store the normalization.
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...