Page 194
Operator‘s Manual FERITSCOPE
®
FMP30
Malfunctions and Messages
Unable to complete cor-
rective calibration.
Cause: Measurement
carried out incorrectly.
Repeat the corrective
calibration and perform
the calibration measure-
ment correctly (e.g., do
not allow the probe to
hover over the standard
before or after the mea-
surement; observe the
minimum lift-off dis-
tance!)
Cause: Calibration stan-
dards used that did not
exhibit the required
-
ferrite content or are de-
fective.
Repeat the corrective
calibration with correct
and good calibration
standards.
Calibration standards
were measured out of
sequence during the cor-
rective calibration (stan-
dard 1 mixed up with
standard 2) and
-
fer-
rite content was not ad-
justed correspondingly
using the arrow keys.
Repeat the corrective
calibration and mea-
sure the standards in
the correct sequence.
Unable to complete the
master calibration.
Cause: Measurement
carried out incorrectly.
Repeat the master cali-
bration and perform the
calibration measure-
ment correctly (e.g., do
not allow the probe to
hover over the standard
before or after the mea-
surement; observe the
minimum lift-off dis-
tance!)
Malfunction/Message Cause
Correction
start
on
Page
E
011
Measurements
out
E
012
Invalid std.
sequence !
E
013
Countrate out
of interval !
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...