Page 110
Operator‘s Manual FERITSCOPE
®
FMP30
Measuring
7.8
Measurements in Standard and Matrix Measuring Mode
The preparations required for making the actual measurement are indepen-
dent of the selected measuring mode and are described in the Chapters
“
7.1 ‘Preparing for a Measurement’, beginning on Page 80 and
7.3 ‘Making a Measurement’, beginning on Page 85.
Measuring modes can be changed from the service functions only
(
10.6.3 ‘Matrix Mode’, beginning on Page 169).
7.8.1
The Standard Measuring Mode
In the Standard measuring mode, measurements are made on the same
specimen (e.g., on a bolt) in succession and are combined to a block and
evaluated using
BLOCK-RES
. The resulting block mean value represents the
local
-ferrite content of the reference area.
In the standard measuring mode, readings can be stored only in the respec-
tive latest not yet closed block of an Application.
The individual applications may contain different numbers of blocks and
different numbers of readings may be stored in a block. However, if auto-
matic block creation is enabled in an Application, only blocks of the set
block size can be created (
5.6.3 ‘Automatic Block Size and Block Cre-
ation’, beginning on Page 49 and
7.3.6 ‘Measurements With a Fixed
Block Size’, beginning on Page 91).
The instrument is re-initialized automatically when changing from the stan-
dard to the matrix measuring mode. During a re-initialization, all Applica-
tions and all measurement data stored in them are deleted and all settings
of the service functions are reset to their default settings. After the re-initial-
ization, i.e., also after changing the measuring mode, the required Appli-
cations will need to be set up anew and the desired settings need to be
established again in the service functions!
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...