Page 96
Operator‘s Manual FERITSCOPE
®
FMP30
Measuring
Procedure - Checking the Current Calibration State
Key sequ./
Action
Detail of the display Explanation
CAL +
Use
CAL
and
to start the intrument fucti-
on CAL-Check.
or
Use the arrow keys to set the rated
-ferrite
content value of the used standard.
ENTER
Use
ENTER
to confirm the entry.
or
Select the form in which the specification of
the measurement uncertainty for the used
standard is available.
Tolerance [
Unit
]: A tolerance is specified for
the used standard.
Example: 2.66 FN ± 0.4 FN
rel. Tolerance [%]
: A percentage tolerance
is specified for the standard.
Example: 2.66 FN ± 5 %
U(k=2) [
Unit
]
: An extended measurement
uncertainty is specified for the used stan-
dard, for an extension factor k = 2. Exam-
ple: 2.66 FN ± 1,2 FN
ENTER
Use
ENTER
to confirm the entry.