Page 76
Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
To verify a correct master calibration, perform several reference mea-
surements.
Now, it is possible to make measurements in the open Application. Delete
the readings of the reference measurement before starting with the mea-
surements on your specimens.
6.4.4
Displaying Xn Ranges for Calibration Standards for the Master
Calibration
During the master calibration, the master characteristic can be determined
only if suitable calibration standards are used.
The calibration standards are suited for the master calibration only if their
normalized countrates Xn are within pre-established Xn ranges. They can
be queried during the master calibration without affecting the calibration.
Please note the information regarding the number of calibration standards
under
Key sequ. /
Action
Detail of the display Explanation
5 x
ENTER
2 x
Retrieve the master calibration from the Ser-
vice functions menu and use
ENTER
to start
the procedure. Relevant explanations
beginning on
BLOCK-RES
Use
BLOCK-RES
to display the Xn range
for calibration standard #1.
[0.0300<Xn<0.2500]
: The normalized
countrates Xn of calibration standard #1
must be within these limits of the Xn range.
BLOCK-RES
Use
BLOCK-RES
to display the Xn range
for calibration standard #2
[0.3000<Xn<0.7000]
.
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...