Operator‘s Manual FERITSCOPE
®
FMP30
Page 97
Measuring
or
Use the arrow keys to set the uncertainty of
the standard. Example: rel. Tolerance
ATTENTION
: Observe the display unit when
entering the uncertainty.
ENTER
Use
ENTER
to confirm the entry.
Standard
Perform a check measurement on standard
(1). We recommend carrying out 10 or
more measurements on the standard (=CAL-
rated value).
[Delete: DEL]
: Use
DEL
to delete the last mea-
surement, 2x
DEL
to delete all readings ob-
tained for the control measurement.
[
Cancel: CAL
]: Using
CAL
cancels the CAL-
check function.
ENTER
Use
ENTER
to end the control measurement.
The result of the check appears in the dis-
play.
[
OK: ENTER
]: Use
ENTER
to finish the CAL-
Check function and return to the measuring
mode. The instrument is ready for measure-
ments.
[
Info: FINAL RES
]: Use
FINAL RES
to open
the test result display of the check measure-
ment.
[
Remeasure: DEL
]: Use
DEL
to repeat the
CAL-check for the same or another stan-
dard.
Key sequ./
Action
Detail of the display Explanation
„
CAL is not OK!
“
Carry out a new corrective calibration,
as the measured mean value of the con-
trol measurement has an excessive error
to the nominal value of the standard.
Use
Cal
to perform a new corrective cal-
ibration.
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...