![Fisher FERITSCOPE FMP30 Скачать руководство пользователя страница 108](http://html1.mh-extra.com/html/fisher/feritscope-fmp30/feritscope-fmp30_operators-manual_2294714108.webp)
Operator‘s Manual FERITSCOPE
®
FMP30
Page 103
Measuring
7.6.4
Overwriting Individual Erroneous Measurements at a Later Time
During the evaluation of the current block, individual erroneous measure-
ments of the current block or of earlier blocks can be overwritten with new
readings.
The instrument does not carry out outlier rejection when overwriting stored
measurement data (even if outlier rejection is enabled).
Key sequ. /
Action
Detail of the display Explanation
BLOCK-RES
Use
BLOCK-RES
to start the evaluation of
the open block (
Current Block “Block Result”’, beginning on
Page 118).
[Blck:]
: Number of the current block
[n=]
: Number of readings stored in the cur-
rent block.
DATE
and
TIME
display
[fe.]
: Displays the mean value of the current
block.
[s]
: Standard deviation of the measure-
ments made
[COV]
:
Coefficient of variation
or
Use the arrow key
or
to select the
block reading that is to be overwritten is
located.
identifies the closed block.
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...