Page 182
Operator‘s Manual FERITSCOPE
®
FMP30
Instrument Settings - Service Function Menu
since this last manual air value capture.
When becomes a static air value measurement necessary?
If the probe contacts the specimen with very different speeds, or if the
touch-down procedure is very slow.
If the probe cannot be lifted off sufficiently from the specimen between
the measurements, e.g., when measurements are made in pipes or
bore holes with inside probes.
Switching on/off the air value acceptance
To keep the measurement errors with static air value measurements as
low as possible, manual air value capture should be carried out at
relatively regular intervals.
Key sequ. /
Action
Detail of the display Explanation
Select
Measurement
by pressing the arrow
key
or
and confirm the selection with
ENTER
.
ENTER
Select
Air value acceptance
by pressing
the arrow key
or
and confirm the selec-
tion with
ENTER
.
ENTER
Select the desired acceptance by pressing
the arrow key
or
.
The settings of the air value acceptance of
the other Applications will not be affected.
[dynamic]
: The air value is captured con-
tinuously during measurement.
[static]
: No air value is captured during
measurement. The air value should be cap-
tured manually in regular intervals, see
Page 183.
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...