Page 66
Operator‘s Manual FERITSCOPE
®
FMP30
Normalization, Calibration and Master Calibration
ENTER
[Entry:
]
: Use the arrow keys to set the
rated
-
ferrite content value of the calibra-
tion standard.
The rated value can be set faster if a mea-
surement is performed on a calibration stan-
dard that is within the tolerance range, and
then the rated value is corrected using the
arrow keys.
[CAL-rat. 1: 0.680]
: Display of the set rat-
ed value for the
-
ferrite content of the cali-
bration standard (Example: 0.680 Fe%)
[Cancel: ENTER]
: Using ENTER cancels
the calibration procedure
Calibration
standard
Make about 5 measurements at different lo-
cations of the calibration standard.
Displayed is the mean value of all measure-
ments performed for this step.
[Delete: DEL]
: Use
DEL
to delete the last
measurement, 2x
DEL
to delete all read-
ings obtained for the calibration.
[OK: ENTER]
: Use
ENTER
to end the cur-
rent calibration step
.
ENTER
If a corrective calibration is desired with 2
or 3 standards, proceed in the same man-
ner with calibration standards 2 and 3.
Otherwise: Use
ENTER
to end the correc-
tive calibration
.
The new characteristic will be computed au-
tomatically and stored.
The instrument is again ready to make mea-
surements.
Key sequ. /
Action
Detail of the display Explanation
Содержание FERITSCOPE FMP30
Страница 1: ...FERITSCOPE FMP30 Operator s Manual Coating Thickness Material Analysis Microhardness Material Testing ...
Страница 63: ...Page 58 Operator s Manual FERITSCOPE FMP30 Applications Fig 5 3 Example for linking of applications ...
Страница 235: ...Page 230 Operator s Manual FERITSCOPE FMP30 Glossary ...
Страница 240: ...Operator s Manual FERITSCOPE FMP30 Page 235 X Xn 79 ...
Страница 241: ...Coating Thickness Material Analysis Microhardness Material Testing ...